The Electrolytic Polisher is used in restorative dentistry. It can be used to polish objects previously treated with a sandblaster, such as cobalt-chrome alloys or stainless steel items including denture frames, crowns, bridges, and clasps.
The two cylindrical electrolytic baths can operate simultaneously. The unit supports multi-directional polishing. The electrolytic baths serve as the cathode, and the clip used to hold the casting serves as the anode. The timer controls the duration of electrolytic polishing.
Specifications:
Voltage: 110V / 220V, 50Hz ±10%
Power: 150W
Timer: 0–15 min
Working current: 10A
Weight: 11.5 kg
Instructions:
1. After unpacking, check that all accessories are included. Read this manual carefully before use.
2. Connect the machine to an electrical outlet and turn on the power. The power indicator should light up, and the machine enters standby mode.
3. Pour the appropriate amount of electrolyte into the baths.
4. Tie the casting to a metal wire and hang it on the L-shaped bracket. Soak the casting completely in the electrolyte within the cathode ring. Do not let the casting on the anode touch the cathode, to avoid short circuit.
5. Adjust the timer clockwise to the set time and turn on the power switch.
6. Turn the current knob to control the electrolytic current. The current amount is indicated on the ammeter. Adjust the current to 4.5–5A for double bath operations.
7. The machine will stop automatically after the timer runs out (4–6 minutes). To stop the machine immediately, turn the power knob anti-clockwise to 0. For polishing small objects, use a low current and monitor the process to reduce the risk of over-polishing.
Cautions:
1. Make sure the power supply voltage matches the voltage of this machine.
2. Check the electrolytic bath regularly and replace it when there is a leak.
3. Make sure there is good contact among the polished casting, stainless steel wire, and clip.
4. Check for short circuit among the polished item, stainless steel wire, and cathode. Make sure there is no short circuit before use.
5. The surface of the polished casting must be treated properly (deoxidizing, degreasing, etc.), otherwise the result may be affected.
6. Make sure the temperature of the electrolyte does not exceed the temperature rating (50°C).
7. Change the electrolyte regularly if the electric current decreases or stays the same after the voltage has been raised. If the electrolytic process takes longer than usual, the electrolyte may need to be changed.
8. Stop the machine for 30 minutes after 90 minutes of continuous use.
9. Make repairs only after the power is off.
10. There must be good contact between the cathode pole and the cathode holes.
11. There should be a reliable ground connection to the chassis of the machine.
Packing List:
Machine × 1


Contact Information:
Email: service.alandental@gmail.com
If you have any problem, please do not hesitate to contact us.
Shipment Note:
If your package shipped by Express Delivery, please give us a valid delivery phone number, and also a full and accurate shipping address.
How to Buy from Alandental?

Email: service.alandental@gmail.com
TEL:+86-023-68438829
If the phone can not be connected for the moment?
Australia Online Dental Equipment/Supplies For Sale
Copyright 2021 www.alandental.com. All rights reserved.