Product name: Dental Ultrasonic Scaler YS-CS-A(5E)
Item no.: YS-CS-A(5E)
Rated input: 48VA
Voltage: AC100–240V 50–60Hz
Ultrasonic wave output characteristics: Maximum output intensity 8W
Transformer input: AC100–240V 50–60Hz; Output: DC 24V
Main vibration excursion: 0.05–0.15mm; Half excursion power: ≥0.5N
External charger: Input: 110V/50–60Hz or 220V/50–60Hz; Output: 5V DC
Tips output vibration frequency: 25kHz–35kHz
Functions: Scaling / periodontal / root canal / polish
Environment temperature: 5°C–40°C
Relative humidity: ≤80%
Atmospheric pressure: 86kPa–106kPa
1. Scaler Function
1.1 Press the power button, pick up the scaler handpiece, and press the "Scaler & Polish" button. Press the "Option" button to select the function until the "G" indicator light is on.
1.2 Press the "Water Mode" button to select the water mode: bottle water supply or external water supply.
1.3 Select the correct working tip and install it on the handpiece using the torque wrench.
1.4 Press the foot pedal. The working tip vibrates, the LED light turns on, and cold water sprays out.
1.5 Vibration strength: Adjust the vibration strength as needed. The default gear is 5.
1.6 Water volume: If the water mode is bottle water supply, adjust the water button for bottle supply. If the water mode is external water supply, adjust the water button for external supply.
1.7 After finishing work, allow water to spray for 30 seconds to rinse the handpiece and working tip, then disinfect the working tip.
2. Periodontal Function
2.1 Install the periodontal working tip using the torque wrench. Press the "Option" button to select the function until the "P" indicator light is on.
2.2 Refer to the scaler function for other steps.
3. Root Canal / Endodontic Function
3.1 Install the root canal working tip using the torque wrench. Press the "Option" button to select the function until the "E" indicator light is on.
3.2 The default gear for root canal is 1. Use the "+" and "-" buttons to select the suitable gear.
3.3 Select the suitable working tip and insert it into the root canal, then press the foot pedal.
3.4 Suggested gear: 1–5 for root canal function.
4. Powder Polish Function
4.1 Add powder into the powder bottle, above "Min" and below "Max".
4.2 Pick up the polish handpiece and press the "Scaler & Polish" button to switch to polish mode.
4.3 Adjust the "Water" and "Air" buttons, then press the foot pedal to confirm water and powder output.
4.4 Wear goggles and a mask for the patient. The user should also wear goggles before treatment.
4.5 Adjust the water and air to the suitable gear. Suggested water gear starts from 5; air gear starts from 3. Adjust according to the patient's teeth status.
4.6 During treatment, keep the working tip about 5mm from the teeth and maintain a beam angle of 30°–60°. Do not let the working tip touch the teeth directly.
4.7 After treatment, adjust the water to the maximum gear and polish all the teeth.