The Way to Replace Parts of Dental Sterilizer

Autoclave sterilizers have been known to play a significant role in the daily functioning of dental operatories and are recently becoming even more advanced and efficient due to advances in technology. Development of a new generation of autoclaves within the last two decades has added a new dimension to this heat sterilization modality. These autoclaves are classified as “Class B” sterilizers or “pre- and post-vacuum” steam sterilizers. The equipment is fitted with a pump that creates an initial vacuum in the chamber to ensure air is removed from the sterilizing chamber before steam enters.

The latest innovations allow for fast and easy use for efficient instrument processing, reliability and effectiveness for infection control assurance, automatic opening door for efficient instrument drying, and maximum space utilization and a complimentary design for your sterilization area. Although these sterilizers can be a somewhat costly addition to a dental operatory, the time and money that they save dentists down the road is substantial. Taking a few pointers into consideration and performing routine checks and replacing old parts on your unit can help maximize the life and functioning of your model.

Door Gasket: To replace the door gasket, begin by simply removing the old door gasket. Clean the gasket groove in the door, and chamber face. For easy installation of the new gasket, use the “South-East-West-North” method (i.e. begin installing the gasket at the bottom of the door, then install the left and right sides and finishing at the top by squeezing in any of the excess making sure that the gasket fits flush).

Dam Gasket: Use CAUTION when removing or installing the dam gasket –the door studs can be sharp. To install the door dam, align and place the dam gasket on and over the door studs. The dam must lie flat against the door (refer to the Installation Instructions included with the MIG035 dam gasket).

Chamber & Vent Filters (mesh style filter): If the machine has both a Chamber and a Vent Filter – Remove all trays and the rack from the chamber. Clean the chamber, then remove the chamber filter. Install the new filter. You must also replace the Vent Filter that serves as a filter for the steam trap/bellows and is located in the vent line to the tee fitting. Reinstall rack and trays.

Sterilizer: After replacing the PM parts, check the reservoir and clean if dirt or debris are present (Note: Debris is the number one reason for Steam Trap/Bellows failure). Refill reservoir with distilled water only. Conduct a temperature check using a Max Register Thermometer. Conduct a pressure check to identify any leaks that may need to be addressed. If calibration is necessary, use the RPI Pressure Gauge and the RPI Trim Pot Tool.